17 Technical specifications
Ambient conditions
During operation:
• Temperature
• Air pressure
• Relative humidity
10 to 40°C (50 to 104 °F)
70 to 160 hPa
10 to 90 %, non-condensing
During transportation:
• Temperature
• Air Pressure
• Relative Humidity
-20 to 60°C (-4 to 140 °F)
50 to 106 kPa
10 to 95%, non-condensing
During storage:
• Temperature
• Air pressure
• Relative humidity
0 to 40°C (+32°F to 104 °F)
70 to 106 kPa
20 to 80%, non-condensing
IMPORTANT:
When moving the device between ambient and controlled temperatures, removing the
device from storage, and before use; acclimate the fabian device for 12 hours.
The Minimum and Maximum ambient temperatures in the clinical area will NOT affect the
accuracy of the Flow, Pressure or Oxygen measurement.
Monitoring
Acoustic Pressure of Alarm tones. Alarm tone sequence IEC 60601-1-8:
Lowest possible setting:
Highest possible setting:
A weighted sound pressure level in 1m distance:
Expiratory Minute Volume
Alarm Lower threshold Value:
Alarm Upper threshold Value:
If Lower Alarm Limit not reached
If Upper Alarm Limit exceeded
Airway Pressure
Alarm Lower threshold Value:
Alarm Upper threshold Value:
If Lower Alarm Limit not reached
If Upper Alarm Limit exceeded
Inspiratory O
2
concentration
Alarm:
Set Value ±5 Vol% for >1 minute at 21 Vol%
below threshold Value 18 Vol%
If no breathing activity is recognized