63 10 Ventilation parameters
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
10 Ventilation parameters
Backup rate in DUOPAP mode
Continuous Positive Airway Pressure
Continuous flow for producing airway pressure
at the CPAP level
Flow setting
Adjustable from 0 to 19 L ⁄ min
Duration of O
2
Flush*
Adjustable from 0 to 120 seconds.
*In Section “16: Setting ranges and parameters”
Maximum time manual breath*
Maximum duration allowed for manual breaths.
Adjustable from 2 to 30 seconds.
*In Section “16: Setting ranges and parameters”
Oxygen concentration
Inspiratory oxygen concentration setting
Oxygen flush
Perform O
2
flush/oxygen spray
(Preoxygenation for max. 2 minutes)
Upper inspiratory pressure
The lower pressure level is selected in DUOPAP mode from
the CPAP parameter.
Manual Inspiratory pressure
Applying a manual breath
Next breath possible after a Block period of 200 milliseconds.
Trigger
The trigger sensitivity can be adjusted from level 1 to 10 and
can be turned OFF.