33 5 System operation
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
Emergency shutdown
WARNING
:
The emergency shutdown procedure should only be used during an actual emergency
situation. This procedure disconnects all power to the ventilator and any other devices
connected to the ventilator.
Before performing this procedure, ensure that the alternate ventilation method (
for
example
: resuscitation) is present, ready and standing by.
Remove patient form the
ventilator.
Press and hold the
start/stop button (1)
for one to two seconds,
then release.
After releasing, press and hold the
start/stop button (1) again for five
seconds.
After the device has powered OFF, a
notice tone will sound.
Press the Alarm Silence button to confirm system switch OFF.
Hold the button for at least three seconds until the WARNING LED
stops.