6 Configurations menu 42
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
Ventilation parameter settings
Figure 6-10: Ventilation parameter settings menu
#
Parameter
Description
O
2
Flush Time
O2 flush duration (Preoxygenation) configuration:
(OFF/30 /60 /120 sec.)
Automatic O
2
Calibration
21 Vol% / 21 and 100 Vol.%. In cases where oxygen supply source
does NOT provide 100% oxygen, this should be set to 21% only to
avoid calibration at false value.
NIV Leak
Compensation
Set either OFF, Low, Middle or High level of NIV leak
compensation (0 up to 10, 20, 30%) in nCPAP and DuoPAP
modes.
Unit for Pressure
Selection of Pressure unit (mbar/cmH
2
O).
NIV Tube Set
Selection of the nCPAP system used.(Infant Flo or InspireTM),
Infant Flow
®
LP, Medijet
®
Maximum time
for Manual
breath
Inspiration time for a manual breath (2 to 30 seconds.)
Hospital
Settings
Optional
: The “Hospital Settings button is only active if special
settings were previously saved in the device.
Hospital Settings can be accessed, loaded from the Service menu
or copied in Service mode.
1
2
3
4
5
6
7
8
9