15 Sterilization / cleaning / disinfection 86
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
15.2.1.1 Preparation for disinfection
1. Switch OFF the fabian before cleaning and unplug at the mains in order to prevent
damage due to penetration by liquid.
2. Remove the Flow sensor from the flow sensor cable.
3. Remove any visible soiling on the flow sensor cable with the cleaning agent using a
manual disinfection option listed in section 15.2.1.2:”Manual disinfection options.
15.2.1.2 Manual disinfection options
Option A: CaviWipes Surface Disinfectant Wipes (EPA REG. NO.: 46781-14)
1. Unfold the CaviWipe and wipe the surface of the enclosure completely.
NOTE
: The surface treated must remain visibly moist for three minutes.
Use more wipes if necessary, to keep the surface moist for three minutes.
2. Wipe the cable with a clean lint-free cloth dampened with purified water for one
minute.
3. Allow to dry in the air.
Option B: 10% bleach (Clorox EPA REG. NO.: 5813-1)
1. Prepare a 10% bleach solution using purified water.
Use a clean lint-free cloth dampened with the bleach solution.
2. Wipe the cable completely.
Remove any heavy soiling with additional dampened cloths if necessary.
NOTE
: The surface treated must remain visibly moist for five minutes.
Use more dampened cloths if necessary, to keep the surface moist for five minutes.
3. Wipe the cable with a clean lint-free cloth dampened with purified water for one
minute.
4. Allow to dry in the air.
Option C: 5% H
2
O
2
, Hydrogen Peroxide (EPA REG. NO.: 335-1)
1. Prepare a 5% H
2
O
2
solution using purified water.
Use a clean lint-free cloth dampened with the H
2
O
2
solution
2. Wipe the cable completely.
Remove any heavy soiling with additional dampened cloths if necessary.
NOTE
: The surface treated must remain visibly moist for five minutes.
Use more dampened cloths if necessary, to keep the surface moist for five minutes.
3. Wipe the cable with a clean lint-free cloth dampened with purified water for one
minute.
4. Allow to dry in the air.