Performance Tests and Calibration 6
U1253B User’s and Service Guide 163
Calibration from Front Panel
Calibration process
The following general procedure is the recommended method
to complete a full instrument calibration.
1 Read and implement “Test Considerations” on page 142.
2 Perform the verification tests (refer to Table 6- 2 on
page 145) to characterize the instrument.
3 Perform the calibration procedures (refer to “Calibration
procedures” on page 164; read also “Adjustment
Considerations” on page 158).
4 Secure the instrument after calibration.
5 Take note of the new security code (if it has been
changed) and the calibration count in the instrument
maintenance records.
Make sure to quit the adjustment mode before switching off the
instrument.