166 U1253B User’s and Service Guide
6 Performance Tests and Calibration
Figure 6-7 Typical calibration process flow
Press
to select calibration item
or
if not performing complete
calibration
Set up calibrator according
to indicated reference value
Press Save when ready
to start calibration
Reading
out of
acceptable
range
Reading is within
acceptable range
Error message appears
for 3 seconds
The word “PASS” appears
for 3 seconds, then proceeds
to the next calibration item