Maintenance 5
U1253B User’s and Service Guide 135
Fuse replacement
Replace any blown fuse in the multimeter according to the
following procedures:
1 Turn the multimeter off and disconnect the test leads.
Ensure that the charging adapter is also removed, if it is
attached to the multimeter.
2 Wear clean and dry gloves on your hands and avoid
touching any components except the fuse(s) and plastic
parts. It is not necessary to recalibrate the multimeter
after replacing a fuse.
3 Remove the battery cover compartment.
4 Loosen two side screws and one lower screw on the
bottom case and remove the bottom case.
5 Loosen the two screws on the top corners to take out the
circuit board.
6 Gently remove the defective fuse by prying one end of the
fuse loose and removing it from the fuse bracket.
7 Replace with a new fuse of the same size and rating.
Make sure the new fuse is centered in the fuse bracket.
8 Ensure that the knob of the rotary switch on the top case
and the corresponding switch on the circuit board remain
at the OFF position.
9 Refasten the circuit board and the bottom cover.
10 Refer to Table 5- 4 on page 136 for the part number,
rating, and size of the fuses.
NOTE
This manual provides only the fuse replacement procedures, but not the
fuse replacement markings.

Table of Contents

Other manuals for Agilent Technologies U1253B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1253B and is the answer not in the manual?

Agilent Technologies U1253B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1253B
CategoryMultimeter
LanguageEnglish

Related product manuals