Fuse replacement
Replace any blown fuse in the multimeter according to the
following procedures:
1 Turn the multimeter off and disconnect the test leads.
Ensure that the charging adapter is also removed, if it is
attached to the multimeter.
2 Wear clean and dry gloves on your hands and avoid
touching any components except the fuse(s) and plastic
parts. It is not necessary to recalibrate the multimeter
after replacing a fuse.
3 Remove the battery cover compartment.
4 Loosen two side screws and one lower screw on the
bottom case and remove the bottom case.
5 Loosen the two screws on the top corners to take out the
circuit board.
6 Gently remove the defective fuse by prying one end of the
fuse loose and removing it from the fuse bracket.
7 Replace with a new fuse of the same size and rating.
Make sure the new fuse is centered in the fuse bracket.
8 Ensure that the knob of the rotary switch on the top case
and the corresponding switch on the circuit board remain
at the OFF position.
9 Refasten the circuit board and the bottom cover.
10 Refer to Table 5- 4 on page 136 for the part number,
rating, and size of the fuses.
This manual provides only the fuse replacement procedures, but not the
fuse replacement markings.