138 U1253B User’s and Service Guide
5Maintenance
Table 5-5 Basic troubleshooting procedures
Malfunction Troubleshooting procedure
No OLED display after switching ON Check battery. Charge or replace battery.
No beeper tone Check the Setup mode to verify whether the beeper function has been set to
OFF. If so, select the desired driving frequency.
Failed to measure current Check the fuse.
No charging indication
[1]
Check 440 mA fuse.
Check external DC adapter to ensure that its output is DC 24 V and that the
plugs are inserted properly into the charging terminals.
Battery life very short after full
charge/Battery not able to charge
after pro-long storage period
Check if the correct rechargeable battery is used.
Try to charge and discharge for 2 or 3 cycles in order to maintain the battery’s
highest capacity.
•NOTE: The performance of the rechargeable battery may degrade with time.
Remote control failure The Agilent logo on the IR-USB cable connected to the multimeter should be
facing up.
Check the baud rate, parity, data bit, and stop bit (default is 9600, None, 8, and
1) in the Setup mode.
Ensure that the required driver for IR-USB has been installed.
Notes for basic troubleshooting procedures table:
1 Never turn the rotary switch of the multimeter from the OFF position when it is charging.

Table of Contents

Other manuals for Agilent Technologies U1253B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1253B and is the answer not in the manual?

Agilent Technologies U1253B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1253B
CategoryMultimeter
LanguageEnglish

Related product manuals