EasyManuals Logo
Home>Agilent Technologies>Multimeter>U1253B

Agilent Technologies U1253B User Manual

Agilent Technologies U1253B
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #213 background imageLoading...
Page #213 background image
Specifications 7
U1253B User’s and Service Guide 185
Table 7-5 Accuracy specifications ± (% of reading + number of LSD) for AC+DC current
Function Range Resolution
Accuracy for AC+DC current
[1][2]
Overload
protection
30 Hz to 45 Hz 45 Hz to 1 kHz
1 kHz to 20
kHz
Current
500.00 µA
[3]
0.01 µA 1.1 + 25 0.8 + 25 0.8 + 25 440 mA
5000.0 µA 0.1 µA 1.1 + 25 0.8 + 25 0.8 + 25 10 × 35 mm
50.000 mA 0.001 mA 1.2 + 25 0.9 + 25 0.9 + 25 AC/DC 1000 V
440.00 mA 0.01 mA 1.2 + 25 0.9 + 25 0.9 + 25
30
kA/fast-acting
5.0000 A 0.0001 A 1.8 + 30
[4]
0.9 + 30
3.3 + 70 < 3A /
5 kHz
11 A
10.000 A 0.001 A 1.8 + 30
[4]
0.9 + 25
3.3 + 70 < 3A /
5 kHz
-
Notes for AC + DC current specifications:
1 Current can be measured from 2.5 A up to 10 A continuously. An additional 0.5% needs to be added to the specified
accuracy if the signal being measured is in the range of 10 A to 20 A for 30 seconds maximum.. After measuring a current
of > 10 A, it is recommended that the meter is left to cool down (in switched OFF state) for twice the length of
measurement time used, before using it again to make low-current measurement.
2 These specifications are defined for signal input > 5% of range.
3 Input current > 35 µArms.
4 Input current < 3 Arms.
5 For 5 A and 10 A ranges, the frequency is verified for less than 5 kHz

Table of Contents

Other manuals for Agilent Technologies U1253B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1253B and is the answer not in the manual?

Agilent Technologies U1253B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1253B
CategoryMultimeter
LanguageEnglish

Related product manuals