166 U1253B User’s and Service Guide
6 Performance Tests and Calibration
Table 6-4 List of calibration items
Function Range Calibration item
[1]
Reference input
AC V
(with rotary switch at
and
[2]
)
5 V 0.3000 V (1 kHz)
3.0000 V (1 kHz)
3.0000 V (10 kHz)
0.3 V, 1 kHz
3 V, 1 kHz
3 V, 10 kHz
50 V 3.000 V (1 kHz)
30.000 V (1 kHz)
30.000 V (10 kHz)
3 V, 1 kHz
30 V, 1 kHz
30 V, 10 kHz
500 V 30.00 V (1 kHz)
300.00 V (1 kHz)
300.00 V (10 kHz)
30 V, 1 kHz
300 V, 1 kHz
300 V, 10 kHz
1000 V 30.0 V (1 kHz)
300.0 V (1 kHz)
300.0 V (10 kHz)
(done for this function; change
rotary switch position or press
to select the next function
that requires calibration)
30 V, 1 kHz
300 V, 1 kHz
300 V, 10 kHz
DC V Short SHORT Dual banana shorting plug with
copper wire
5 V 3.0000 V 3 V
50 V 30.000 V 30 V
500 V 300.00 V 300 V
1000 V 1000.0 V
(done)
1000 V
DC mV Short SHORT Dual banana shorting plug with
copper wire
50 mV 30.000 mV 30 mV
500 mV 300.00 mV 300 mV
1000 mV 1000.0 mV
(done)
1000 mV

Table of Contents

Other manuals for Agilent Technologies U1253B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1253B and is the answer not in the manual?

Agilent Technologies U1253B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1253B
CategoryMultimeter
LanguageEnglish

Related product manuals