EasyManuals Logo
Home>Agilent Technologies>Multimeter>U1253B

Agilent Technologies U1253B User Manual

Agilent Technologies U1253B
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #218 background imageLoading...
Page #218 background image
190 U1253B User’s and Service Guide
7 Specifications
For current measurements
Peak Hold Specifications
Table 7-11 Frequency sensitivity specifications for current measurements
Input range
Minimum sensitivity (RMS sine wave)
20 Hz to 20 kHz
500 µA 100 µA
5000 µA 250 µA
50 mA 10 mA
440 mA 25 mA
5 A 1 A
10 A 2.5 A
Notes for frequency sensitivity and trigger level specifications for current measurements:
1 For maximum input, please refer to AC current measurements.
Table 7-12 Peak hold specifications for dc voltage and current measurements
Signal width Accuracy for DC mV/V/current
Single event > 1 ms 2% + 400 for all ranges
Repetitive > 250 µs 2% + 1000 for all ranges

Table of Contents

Other manuals for Agilent Technologies U1253B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1253B and is the answer not in the manual?

Agilent Technologies U1253B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1253B
CategoryMultimeter
LanguageEnglish

Related product manuals