SIRIUS®
So, after checking again, that you have the correct version of the firmware, press the Upload button, and wait until the
upload process is finished:
Do not interrupt the firmware-upload!
8.3.3 Module Firmware Upgrade
For the module firmware upgrade, follow the same procedure as for the system firmware upgrade.
Select the device DWUSB Sirius - Modules from the top drop-down list
Make sure to select the firmware file for the modules (the file-name starts with SirisuMod_)
Illustration 231: Firmware-upgrade of Modules
Finally click the Upload button and wait until all modules have been updated:
8.3.4 Troubleshooting
Instrument or FWU file is not correct!
When you get this error-message, it means, that the selected
firmware file does not match the selected Device.
In the example screenshot to the right, we have selected the
Sirius® modules firmware file, but the Device is set to
DEWESoft USB .
So you must either change the Device to
DWUSB Sirius - Modules or select a System firmware file.
Page 158/166 www.dewesoft.com Doc-Version: 1.4.2
DEWESoft® DEWESoft® DEWESoft® DEWESoft® DEWESoft® DEWESoft® DEWESoft® DEWESoft® DEWESoft® DEWESoft® DEWESoft® DEWESoft®
measurement innovation measurement innovation measurement innovation measurement innovation measurement innovation measurement innovation measurement innovation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DEWESOFT SIRIUS and is the answer not in the manual?

DEWESOFT SIRIUS Specifications

General IconGeneral
BrandDEWESOFT
ModelSIRIUS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals