EasyManuals Logo

Dini Argeo 3590ET User Manual

Dini Argeo 3590ET
79 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
3590EKR, 3590EXP, 3590EXT, CPWE, CPWET series indicator E-AF09_02_16.03_EN_U
5.5 MESSAGES OF THE INSTRUMENT
While using the indicator, it is possible to incur into the following errors:
5.5.1 INSIDE THE SET-UP OR MENU OR AT START-UP
MESSAGE
DESCRIPTION
Error S X Cell Y
Communication error on scale X (E.G. Range 1), to Platform number Y
Err. 98
Connection error between the indicator and the platforms, after have showed this
error, the instrument, show which platform has problems.
C.Er. 8.03
When carrying out the multirange or multidivisional calibration, the ranges have not
been entered in an increasing manner (RANGE 1 < RANGE 2 < RANGE 3).
C.Er. 36
CALIBRATION ERROR
During the calibration some internal negative points have been calculated:
- the calibration point is less than the zero point.
- the signal is negative (check the connections)
Err.37 / alternately with No.Cal
CALIBRATION ERROR
During the calibration some internal points less than the minimum value have been
calculated:
- the calibration point is equal to the zero point.
- A capacity too high in relation to the division has been set.
Err. 38
NO CALIBRATION
Range not valid: please check that RANGE 1 < RANGE 2 < RANGE 3.
Er 39
NO CALIBRATION
Invalid calibration range number (there is the value 0 or values greater than 3); one
must execute a TECHNICAL DEFAULT (dFLt.t parameter of the SET-UP
environment), if not already executed previously, and carefully carry out the
calibration.
Er 40
The value 0 is in the “range 1” parameter of the calibration; one must execute a
TECHNICAL DEFAULT (dFLt.t parameter of the SET-UP environment) and carefully
carry out the calibration.
Er 41
The value 0 is in the “diV 1” parameter of the calibration; one must execute a
TECHNICAL DEFAULT (dFLt.t parameter of the SET-UP environment) and carefully
carry out the calibration.
Err. 72
Equalization coefficient adjustment not allowed because a too low sample weight has
been used.
Err. 85
Equalization coefficient adjustment not allowed because the instrument has not been
calibrated.
ErPnt
During the acquisition of a calibration point, a null value from the converter has been
read.
hW-Err
HARDWARE ERROR: software not compatible with the installed hardware; the
hardware expansion component, which allows the software to function, is missing.
EXITING SETUP:
SAVE ?
The instrument requests the saving when exiting the set-up; press ENTER to confirm
or C to not continue..
SURE?
The instrument requests a confirmation before proceeding: press ENTER to confirm
or C to not continue.
Error
Generic error. Possible motives can be:
- Division not valid: please check that DIVISION 1 < DIVISION 2 < DIVISION 3;
- An error has taken place in the connection, configuration, equalisation,
calibration of the cells, etc.
EXECUTION
AUTOZERO
“autozero at start-up” function is active (see section 3.4)

Table of Contents

Other manuals for Dini Argeo 3590ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dini Argeo 3590ET and is the answer not in the manual?

Dini Argeo 3590ET Specifications

General IconGeneral
BrandDini Argeo
Model3590ET
CategoryAccessories
LanguageEnglish

Related product manuals