Preventive maintenance: Battery maintenance
5-16 Dash 3000/4000/5000 2000966-542D
About the Cadex SMart Two+ charger
NOTE
The Cadex SMart Two+ charger is no longer available for purchase.
However, if your site has a Cadex SMart Two+ charger, you should
continue to use it to charge and condition the battery.
Battery charger LED indicators
The following is a quick guide which identifies the meaning of the
charger LEDs.
Equipment software requirements
The patient monitor must use Dash 3000/4000 software version 2A or
later.
The Cadex SMart Two+ charger must use software version 1.31 or
later.
LED indicators Illuminated Flashing
RUN Charging in progress. Initializing the battery.
RUN and CONDITION Conditioning in progress.
READY Charging is complete.
READY and CONDITION Conditioning is complete —
pass target.
Conditioning is required.
FAIL Battery fault. Charger fault.
FAIL and CONDITION Conditioning is complete — fail
target.