Functional and electrical safety checks: Functional Checkout procedures
8-20 Dash 3000/4000/5000 2000966-542D
Functional Checkout procedures
Frequency
Qualified personnel must perform the checkout procedures:
Every 12 months after receipt of the device (Preventative
Maintenance). Refer to “Maintenance schedule” on page 5-2 for more
information.
Each time the main enclosure is disassembled or a circuit board is
removed, tested, or replaced (Corrective Maintenance). Refer to
“Recommended checkout” on page 7-52 for more information.
Identify enabled patient parameters and software options
The check out procedures support all enabled patient parameter and
software options. To view the options enabled on the patient monitor, go
to the Main Menu and select MORE MENUS > MONITOR SETUP >
SOFTWARE CONFIGURATION.
The following is a list of all available patient parameter and software
options.
0BP (invasive)
1BP or 2BP (invasive)
3BP or 4BP (invasive)
12SL
ACI-TIPI
HI-RES TRENDS
ETCO
2
(CO
2
) (End-tidal CO
2
)
NETWORK
AVOA Plus
CARDIO-PULMONARY
CARDIAC

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals