Functional and electrical safety checks: Functional Checkout procedures
2000966-542D Dash 3000/4000/5000 8-45
Test pattern (option based on Dash having a local printer)
Using the Trim Knob control, access the Service Mode menu starting
from the Main Menu.
1. Select MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the Trim Knob control to select the day and
month from patient monitor screen with leading zeros. (e.g. July 4 =
0407).
3. Select PRINT (or GRAPH) TEST PATTERN > START.
4. Verify the following:
Fonts
Shading
Triangle Pattern
No missing dots
5. Select PRINT (or GRAPH) TEST PATTERN > STOP.
Graph speed
Using the Trim Knob control, access the Graph Setup menu starting
from the Main Menu.
1. Select MORE MENUS > MONITOR SETUP > PRINT (or
GRAPH) SETUP.
2. Select SPEED:25 (default).
3. Verify that all eight speeds work.
Display test
1. Hold the NBP Auto and the ZERO ALL keys and press the Trim
Knob control at the same time.
2. Release the Trim Knob control immediately.
3. Continue holding the NBP Auto and the ZERO ALL keys.
4. Select “Video Test Screens.”
5. Test all screens:
White Screen
Red Screen
Blue Screen
Green Screen
Vertical Bars

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals