Configuration: Boot Code settings
2000966-542D Dash 3000/4000/5000 4-21
Chinese Language
NOTE
The following two items are only available when the Chinese
language is selected. The factory default is disabled.
Select 3.2s baseline filter coefficent to enable or disable the 3.2
second baseline filter time constant.
Select Minimum Waveform Width Reduction to enable or disable
the size reduction of the LCD waveform line width.
Enable or disable AFIB Identification
NOTE
This feature is already enabled if the Cardiac software option is
purchased.
Select AFIB Identification to enable or disable the atrial fibrillation
detection feature.
To verify the option has been enabled on the patient monitor,
1. Select the following menu options: MORE MENUS > MONITOR
SETUP > SOFTWARE CONFIGURATION.
2. Verify that ATRIAL FIB is listed under “ENABLED SOFTWARE
OPTIONS.”
To change the status of this feature,
1. Activate Boot Code (“Boot Code settings” on page 4-18) and select
SET CONFIGURATION menu option.
2. Select the AFIB Identification menu option.
3. Select enabled to activate this option, or disabled to deactivate this
option.
4. Select Exit and wait for the patient monitor to display the message,
“Writing settings to Flash config mem...”
5. Manually reboot the patient monitor to implement this change.
6. Refer to the operator’s manual to configure the patient monitor to
use this feature.
Enable or disable IntelliRate
IntelliRate heart rate calculation feature is enabled from the factory.
To verify the option has been enabled on the patient monitor,
1. Select the following menu options: MORE MENUS > MONITOR
SETUP > SOFTWARE CONFIGURATION.
2. Verify that ECG INTELLIRATE is listed under “ENABLED
SOFTWARE OPTIONS.”
To change the status of this feature,