Field replaceable units: Remove or replace handle assembly
7-10 Dash 3000/4000/5000 2000966-542D
Remove or replace handle assembly
1. Remove the two screws holding the handle to the patient monitor.
Take care not to strain the cables if your patient monitor has the
alarm light option or the wireless option.
2. If your patient monitor does not have the alarm light option or
wireless option, replace the defective handle assembly or continue to
“Remove or replace display assembly” on page 7-18 to replace other
parts.
3. If your patient monitor has the alarm light option, replace the
defective handle assembly, or continue to remove the display
assembly to route the alarm light cable through the Dash housing (if
it is not already routed) or to replace other parts. Refer to “Remove or
replace display assembly” on page 7-18.
NOTE
For Dash 3000 only: If the replacement handle assembly has an
alarm light, the DAS assembly must be removed to connect the
cable to J1 on the writer board. Refer to “Replace DAS assembly”
on page 7-37 for instructions on removing and reinstalling the
DAS assembly.
537A
967A
alarm light cable

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals