Network disclosure to facilitate network risk management:
C-2 Dash 3000/4000/5000 2000966-542D
Purpose and scope
This disclosure is intended to satisfy the requirements of IEC 60601-
1:2005 clause 14.13 and IEC/ISO 80001-1:2010 clause 3.5 for disclosure
of network related specifications, requirements and residual risks in
order to facilitate the responsible organization’s risk management
activities (e.g. pursuant to 80001-1) for their networks incorporating the
Dash 3000/4000/5000 Patient Monitor
Purpose for the Dash patient monitor connection to network
The Dash patient monitor is intended to be connected to a network in
order to support the following functionality:
Providing real-time patient data (i.e., parameters, waveforms and
alarms) to compatible network devices such as central stations or
other bedside monitors.
Displaying real-time patient data (i.e., parameters, waveforms and
alarms) from compatible network devices such as other bedside
monitors.
Remote configuration (patient admission, alarm settings, etc.) from
compatible network devices.
Dash patient monitor network interface technical specifications
Connection name Mission critical (MC) network port
Physical network connection type IEEE 802.3 10Base-T Ethernet
Speeds and duplex modes supported 10 Mbps half duplex
Default IP address (from factory) IP address - 126.x.x.x (based on last three octets of MAC address)
Subnet mask - 255.0.0.0
IP addressing IPV4 Static
QoS support IP layer DSCP tagging (packets marked as 26)