Functional and electrical safety checks: Functional Checkout procedures
2000966-542D Dash 3000/4000/5000 8-33
Pulse oximetry tests for Nellcor OxiMax SPO2
Equipment
Use the following equipment for these tests:
Nellcor SRC-MAX SpO
2
Simulator (2007650-002).
NOTE
The Nellcor SRC-MAX SPO
2
Simulator is only compatible with
the Nellcor SPO
2
parameter.
Procedures
1. Verify that 2 AA alkaline batteries are installed in the SRC-MAX
Nellcor Pulse Oximetry functional tester.
2. Connect the SRC-MAX Nellcor Pulse Oximetry functional tester to
the 9-pin end of sensor connector from the Dash. Make sure the Dash
is powered.
3. Verify that the IR LED and RED LED drive indicators are both lit on
the SRC-MAX.
4. Verify the SRC-MAX default indicators are set as follows:
HEART RATE = 60 bpm
LIGHT = LOW
%SpO2 = 75
MODULATION = LOW
Allow the Dash a few seconds to obtain a steady reading.
SpO
2
Port.
Nellcor SpO
2
cable.
Nellcor compatible
SpO
2
Simulator.
892A

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals