Functional and electrical safety checks: Functional Checkout procedures
8-54 Dash 3000/4000/5000 2000966-542D
Dash Port 2 docking station test (option)
Electrical safety tests
When the patient monitor is connected to the docking station, perform
the electrical safety tests described in the maintenance section of the
Dash Port 2 Service Manual.
Operation
Complete the Checkout procedures located in the Dash Port 2 Docking
Station Service Manual.
TRAM-rac 2A module housing peripheral device test (option)
Electrical safety tests
Perform the electrical safety tests on the TRAM-rac 2A separate from the
patient monitor.
Refer to the “Electrical safety tests” on page 8-4 and complete the
following tests.
1. Power outlet test
2. Ground (earth) continuity test,
3. Ground (earth) wire leakage tests, and
4. Enclosure leakage current test.
Operation
Complete the TRAM-rac 2A Module Housing Test procedures found in
the maintenance section of the TRAM-rac 2A Module Housing Service
Manual.
ICG Module test (option)
Refer to the “Electrical safety tests” section of the ICG Module Service
Manual.
Operation
Complete the ICG Test found in the Checkout Procedure section of the
ICG Module Service Manual.