Field replaceable units: Replace main unit parts
2000966-542D Dash 3000/4000/5000 7-37
Replace DAS assembly
WARNING
DAS ASSEMBLY — Do not open the DAS assembly as
this breaks the isolation barrier which may result in
patient death or serious injury. The DAS assembly is a
field replaceable unit only. There are NO field repairs or
adjustments for the DAS assembly.
CAUTION
The patient monitor requires software version 5 or later
to operate with the Masimo SET SPO2 and 4 BP
SuperStat DAS upgrade. If the installation of software
version 5 or later is required, GE recommends that you
FIRST install the software. Then, install the Masimo
SET SPO2 and 4 BP SuperStat hardware upgrade.
1. Remove the two screws holding the handle to the patient monitor.
Refer to page 7-10.
2. Remove the display assembly from the main unit. Refer to page 7-18.
3. Remove the four screws anchoring the Data Acquisition System
(DAS) assembly in place.
CAUTION
CAREFULLY remove the DAS assembly so that you do
NOT hit the components on the CPU/battery housing
assembly.
On the Dash 5000 patient monitor the top two screws attach to a
bracket, not the rear housing as shown.