Electromagnetic compatibility (EMC): Electromagnetic compatibility (EMC)
2000966-542D Dash 3000/4000/5000 A-3
Guidance and manufacturer’s declaration – electromagnetic
immunity
The Dash 3000/4000/5000 is intended for use in the electromagnetic
environment specified in the following table. It is the responsibility of the
customer or user to assure that the Dash 3000/4000/5000 is used in such
an environment.
CAUTION
There are technological limitations in the ability of the
monitor to perform as intended when subjected to
interference from electrostatic discharges (ESD) at or
above ±8kV when tested to EN 60601-1-2. ESD air
discharges into the patient cable connectors and the Aux
port may cause the monitor to reboot and recover to its
normal operating mode without operator intervention.
The following guidelines can greatly reduce the potential
for ESD interference prior to connecting any cables to the
monitor:
Discharge any electrostatic charge from your body
prior to connecting any cables to the monitor.
You can discharge electrostatic charge by touching
the exposed metal equipotential terminal on the back
of the monitor. Refer to “Back view” on page 2-4 for
location of the equipotential terminal.
NOTE
ESD will not cause any permanent damage to the monitor.

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals