Installation: Installation overview
3-2 Dash 3000/4000/5000 2000966-542D
Installation overview
The following illustrates the installation steps for patient-ready
monitoring.
1. Check received equipment against the sales order.
2. Unpack equipment and inventory contents.
3. Inspect the equipment and its components carefully prior to
installation. Refer to “Inspection” on page 3-3.
4. (Optional) Insert two batteries into the battery compartment slots.
(Batteries should be charged for 4 hours before using device on
patients.)
5. Connect the power cord to the power supply inlet on the Dash patient
monitor. Secure the power cord to the back of the Dash using the
power cord retention clamp, unless the Dash is used with the Dash
Port 2 docking station. Then plug the power cord into the AC power
outlet.
6. Connect peripheral equipment and network, if applicable.
7. Power up the Dash monitor. Go to “Configuration” on page 4-1 to
configure for stand-alone or networked monitoring.
8. Connect patient cables per the operator’s manual.
9. Qualified personnel must perform the “Dash installation checkout
procedure” on page 3-8, and should record the values of each required
electrical safety test in the “Checkliston page D-1.
599A

Table of Contents

Other manuals for GE Dash 5000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 5000 and is the answer not in the manual?

GE Dash 5000 Specifications

General IconGeneral
BrandGE
ModelDash 5000
CategoryMedical Equipment
LanguageEnglish

Related product manuals