Functional and electrical safety checks: Overview
2000966-542D Dash 3000/4000/5000 8-3
Functional Checkout procedures
1. “Patient monitor power-up tests” on page 8-21
2. “ECG tests” on page 8-22
3. “Respiration tests” on page 8-25
4. “Temperature tests” on page 8-26
5. “Cardiac output tests (option)” on page 8-27
6. “Invasive blood pressure tests (option)” on page 8-27
7. “Pulse oximetry tests for GE Ohmeda SPO2 oximeter” on page 8-30
8. “Pulse oximetry tests for Masimo SET SPO2” on page 8-32
9. “Pulse oximetry tests for Nellcor OxiMax SPO2” on page 8-33
10. “Noninvasive blood pressure tests” on page 8-35
11. “NBP calibration” on page 8-37
12. “Analog output and defibrillator synchronization tests” on page 8-40
13. “End-tidal CO2 test (option)” on page 8-44
14. “Battery tests” on page 8-44
15. “Graph or print tests (option)” on page 8-44
16. “Display test” on page 8-45
17. “Speaker test” on page 8-46
18. “Network test (option)” on page 8-46
19. “Remote control test (option)” on page 8-46
20. “BISx test (option)” on page 8-47
21. “Wireless LAN test (option)” on page 8-52
22. “Dash Port 2 docking station test (option)” on page 8-54
23. “TRAM-rac 2A module housing peripheral device test (option)” on
page 8-54