Field replaceable units: Replace display flex assembly
2000966-542D Dash 3000/4000/5000 7-25
6. Carefully bend the display flex as shown.
7. Pick up and support the display assembly between the main unit and
a thick book.
8. Use two screws to carefully secure the flex processor cable connector
to the main processor PCB.
9. Carefully align and insert the DAS cable pins into the DAS assembly
until the snap latch clicks into place.
10. Reassemble the patient monitor in reverse order of disassembly.
11. Complete the procedures in “Recommended checkout” on page 7-52.
874A
875A
2 screws
876A