Field replaceable units: Recommended checkout
7-54 Dash 3000/4000/5000 2000966-542D
NBP assembly “NBP
calibration” on
page 8-37.
“Noninvasive blood pressure tests” on page 8-
35.
“Electrical safety tests” on page 8-4.
Wireless LAN None “Patient monitor power-up tests” on page 8-21.
“Network test (option)” on page 8-46.
“Wireless antenna signal strength test (option)”
on page 8-49.
“Wireless LAN test (option)” on page 8-52.
Wireless card None “Wireless antenna signal strength test (option)”
on page 8-49.
“Wireless LAN test (option)” on page 8-52.
Power supply
assembly
None “Patient monitor power-up tests” on page 8-21.
“Wireless antenna signal strength test (option)”
on page 8-49. Use this procedure if the wireless
option is installed.
“Wireless LAN test (option)” on page 8-52. Use
this procedure if the wireless option is installed.
CPU “End-tidal
CO2 test
(option)” on
page 8-44.
“NBP
calibration” on
page 8-37.
“Patient monitor power-up tests” on page 8-21.
“Analog output and defibrillator synchronization
tests” on page 8-40.
“Graph or print tests (option)” on page 8-44.
“Display test” on page 8-45.
“Speaker test” on page 8-46.
“Network test (option)” on page 8-46.
“Wireless antenna signal strength test (option)”
on page 8-49. Use this procedure if the wireless
option is installed.
“Wireless LAN test (option)” on page 8-52. Use
this procedure if the wireless option is installed.
“TRAM-rac 2A module housing peripheral
device test (option)” on page 8-54.
Writer, writer flex None “Patient monitor power-up tests” on page 8-21.
“Graph or print tests (option)” on page 8-44.
“Wireless antenna signal strength test (option)”
on page 8-49. Use this procedure if the wireless
option is installed.
“Wireless LAN test (option)” on page 8-52. Use
this procedure if the wireless option is installed.
Recommended calibration, electrical safety tests, and checkout procedures
Replacement
procedure
Calibration
tests
Checkout procedures Electrical safety tests