Functional and electrical safety checks: Functional Checkout procedures
2000966-542D Dash 3000/4000/5000 8-39
NBP parameter box) on the patient monitor and manometer are
equal (± 1 mmHg) for at least one full minute.
8. Select CHECK CAL IN PROGRESS > STOP. The patient monitor
automatically releases pneumatic pressure in the entire plumbing
circuit.
0
9. Unplug the patient monitor from AC power source and remove the
test apparatus from the patient monitor.