IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-19
PRELIMINARY
7-8-4 BEP Fan does not Spin (ON/OFF button is Orange)-->MPB to BIB Connection
Symptom: T-CFE and T-FEPS LEDs turn On and then off; BEP fan does not spin,
The On/Off button on the cockpit turns green for a second and then turns orange)
Suggested solution: Reset MPB to BIB connection
1.) Verify BEP jumper is in place
2.) Reseat MPB to BIB cable
Figure 7-40 MPB to BIB cable Reseat
3.) Check BIB connector for bent pins
4.) Replace BIB board
Figure 7-39 T-CFE LEDs; T-FEPs LEDs