8-8-2 Wi-Fi Adapter Replacement/Installation Procedure
8-8-2-1 Tools
Appropriate Phillips screwdriver
8-8-2-2 Time Required
15 min
8-8-2-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-8-2-4 Wi-Fi Adapter Removal Procedure
NOTE: Perform these steps only for Venue™ systems with Wi-Fi adapter already installed on them.
For systems with no previously installed Wi-Fi adapter skip this section and proceed to section
Wi-Fi Adapter Installation Procedure (For Systems with No Existing Wi-Fi Adapter)
1.) On the back of the cockpit, release four screws and remove the transparent Wi-Fi dongle cover.
2.) Disconnect the Wi-Fi dongle.
FRU Part #
Refer to Table 9-19 on page 9-18.
Figure 8-181 Wi-Fi Cover and Dongle Removal