8-80 Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-6-2 MPB Guide L and Guide R Replacement Procedure
NOTE: The procedure below is the same whether replacing MPB Guide L or MPB Guide R that are mirrored items
attached to the Receptacle Box side walls.
8-6-2-1 Tools
Appropriate Phillips screwdriver.
8-6-2-2 Time Required
15 min
8-6-2-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-2-4 MPB Guide L/R Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe cover, Lower Front eTower, Side Covers, MPB Power Door
3.) Remove the MPB Module.
4.) Unscrew the two mounting screws that hold the Guide L/R.
FRU Part #
Refer to Table 9-5 on page 9-4.
Accessories - Replacement Procedures
RS Probe Cover Removal Procedure
Lower Front eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
MPB Door Cover Removal Procedure
MPB Module Removal Procedure