IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
Chapter 1 - Introduction 1-13
PRELIMINARY
1-4-4 Electrical Safety
To minimize shock hazard, the equipment must be connected to a well grounded power source. The
system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with safety grounding.
To ensure proper grounding, connect this equipment to a receptacle marked "HOSPITAL ONLY" OR
"HOSPITAL GRADE".
The power outlet used for this equipment should not be shared with other types of equipment. Both the
system power cable and the power connector must meet international electrical standards.
WARNINGWARNING
CONNECTING A VENUE™ SCANNER TO INCORRECT POWER LEVEL WILL DESTROY
THE SYSTEM!!
CONNECT THE SYSTEM ONLY IN ACCORDANCE WITH THE VOLTAGE INDICATED ON
THE PRODUCT LABEL.
WARNINGWARNING
SAFE PRACTICES:
FOLLOW THESE GUIDELINES TO MINIMIZE SHOCK HAZARDS WHENEVER YOU ARE
USING THE ULTRASOUND SYSTEM:
TO MINIMIZE SHOCK HAZARD, THE EQUIPMENT CHASSIS MUST BE
CONNECTED TO AN ELECTRICAL GROUND.
THE ULTRASOUND SYSTEM IS EQUIPPED WITH A THREE-CONDUCTOR AC
POWER CABLE. THIS MUST BE PLUGGED INTO AN APPROVED
ELECTRICAL OUTLET WITH SAFETY GROUND.
THE POWER OUTLET USED FOR THIS EQUIPMENT SHOULD NOT BE
SHARED WITH OTHER TYPES OF EQUIPMENT.
BOTH THE ULTRASOUND SYSTEM POWER CABLE AND THE POWER
CONNECTOR MUST MEET INTERNATIONAL ELECTRICAL STANDARDS.