Chapter 8 - Replacement Procedures 8-147
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-7-9 Printer USB Cable Replacement Procedure
8-7-9-1 Tools
Appropriate Phillips screwdriver
8-7-9-2 Time Required
5 min
8-7-9-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-7-9-4 Printer USB Cable Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe Cover, Lower Front eTower, Left side eTower, Right Side
eTower, Upper eTower Front Cover.
3.) Remove the MPB Front Metal Door
Refer to Table 9-19 on page 9-18.
• Accessories - Replacement Procedures
• RS Probe Cover Removal Procedure
• Upper eTower Front Cover Removal Procedure
• Lower Front eTower Cover Removal Procedure
• Left Side eTower Cover Removal Procedure
• Right Side eTower Cover Removal Procedure
• MPB Front Metal Door Removal Procedure
Figure 8-149 Printer Insert Cover Removed