8-36 Section 8-4 - System Modules- Replacement Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-4-5 Cockpit (Monitor) Replacement Procedure
8-4-5-1 Tools
Appropriate Phillips screwdriver
Flat-head screwdriver
8-4-5-2 Time Required
15 min
8-4-5-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-5-4 Cockpit (Monitor) Removal Procedure
1.) On the back of the cockpit, release four screws and remove the transparent Wi-Fi dongle cover.
Disconnect the Wi-Fi dongle.
2.) Remove four Phillips screws securing the cockpit (monitor) to the arm.
FRU Part #
Refer to Table 9-11 on page 9-11.
Figure 8-22 Wi-Fi Cover and Dongle Removal