Chapter 8 - Replacement Procedures 8-83
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-6-3 Halo Handle Replacement Procedure
8-6-3-1 Tools
LOCTITE® 243™
Allen key 3 mm.
8-6-3-2 Time Required
30 min
8-6-3-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-3-4 Halo Handle Removal Procedure
1.) Remove all accessories.
2.) Remove cockpit (monitor).
3.) Remove the arm.
4.) Remove the Riser Thermal Cover.
5.) Remove the Riser Cover.
FRU Part #
Refer to Table 9-6 on page 9-5.
Accessories - Replacement Procedures
Cockpit (Monitor) Removal Procedure
Arm Removal Procedure
Riser Thermal Cover Removal Procedure
Riser Cover Removal Procedure