8-86 Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-6-4 Articulated Arm Replacement Procedure
8-6-4-1 Tools
Allen key 2.5 mm.
Snap ring (circlip) pliers
8-6-4-2 Time Required
10 min
8-6-4-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-4-4 Arm Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: RS Probe Cover, Lower Front eTower cover, Left Side eTower cover,
Right Side eTower cover and Riser Thermal Cover.
3.) Remove cockpit (monitor).
4.) Carefully raise the arm to its up position.
FRU Part #
Refer to Table 9-6 on page 9-5.
Accessories - Replacement Procedures
RS Probe Cover Removal Procedure
Lower Front eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
Riser Thermal Cover Removal Procedure
Cockpit (Monitor) Removal Procedure