8-50 Section 8-4 - System Modules- Replacement Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-4-9 BEP PCB Replacement Procedure
8-4-9-1 Tools
• Appropriate Phillips screwdriver
8-4-9-2 Time Required
15 min
8-4-9-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-9-4 BEP PCB Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: Lower Front eTower, Left side eTower, and Right Side eTower,
3.) Perform BE module removal procedure.
4.) Remove omega bracket: release 5 screws connecting the omega bracket:
5.) Remove the supporting bracket using an appropriate Phillips screwdriver:
• Accessories - Replacement Procedures
• RS Probe Cover Removal Procedure
• Lower Front eTower Cover Removal Procedure
• Left Side eTower Cover Removal Procedure
• Right Side eTower Cover Removal Procedure
• MPB Front Metal Door Removal Procedure
• BE Module Removal Procedure
Figure 8-36 BE Omega Bracket Removal