Chapter 8 - Replacement Procedures 8-45
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-4-8 Back End (BE) Module Replacement Procedure
8-4-8-1 Tools
Appropriate Phillips screwdriver.
8-4-8-2 Time Required
15 min
8-4-8-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-8-4 BE Module Removal Procedure
1) Remove all accessories.
2) Remove the following covers: Lower Front eTower cover, Left Side eTower cover, and Right Side
eTower cover.
3) Remove the SSD module.
4) Remove MPB Front Metal Door.
Refer to Table 9-13 on page 9-13.
• Accessories - Replacement Procedures
• RS Probe Cover Removal Procedure
• Lower Front eTower Cover Removal Procedure
• Right Side eTower Cover Removal Procedure
• Left Side eTower Cover Removal Procedure
• SSD Module Removal Procedure
• MPB Front Metal Door Removal Procedure