Chapter 8 - Replacement Procedures 8-45
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
8-4-8 Back End (BE) Module Replacement Procedure
8-4-8-1 Tools
Appropriate Phillips screwdriver.
8-4-8-2 Time Required
15 min
8-4-8-3 Preparations
1.) Shut down the Venue™ ultrasound unit, as described in Power Shut Down .
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-4-8-4 BE Module Removal Procedure
1) Remove all accessories.
2) Remove the following covers: Lower Front eTower cover, Left Side eTower cover, and Right Side
eTower cover.
3) Remove the SSD module.
4) Remove MPB Front Metal Door.
FRU Part #
Refer to Table 9-13 on page 9-13.
Accessories - Replacement Procedures
RS Probe Cover Removal Procedure
Lower Front eTower Cover Removal Procedure
Right Side eTower Cover Removal Procedure
Left Side eTower Cover Removal Procedure
SSD Module Removal Procedure
MPB Front Metal Door Removal Procedure