8-206 Section 8-9 - Operating System and Application Software Loading Procedures
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
2.) Perform a full back-up for Patient Archive and System Configuration
(for instructions, refer to the Venue™ User Manual on your Document Media).
3.) Write down the following settings as it will be necessary to restore them at the end of the process:
- TCP/IP address
- Network printer (if present)
- Wireless network settings (if present)
- InSite ExC settings
4.) When done, continue to Tabl e 8-4 on page 8-216.
8-9-4-2 Operating System and/or Application Software Installation Procedure
NOTE: A minimum of 1 hour is required to install both the Windows operating system and the Venue™
application software. This excludes preparation time, backup etc. and performing functional checks
following the installation procedures.
Throughout the installation processes, the Venue™ system will be unavailable for scanning.
Section Section 8-9 on page 8-184 provides step-by-step instructions for installation of Windows
operating system software and Venue™ ultrasound scanner software application. However, these
procedures may be performed separately, as required.
Follow the installation instructions below, as appropriate:
• Windows operating system and Venue™ s/w application:
• Windows operating system ONLY:
• Venue™ software application ONLY:
NOTE: It is necessary to perform the recommended Functionality Checks after completion of the installation
procedures, when installing the following:
• Windows operating system and Venue™ s/w application
• Venue™ software application ONLY