14-4 SCPI Reference Tables
Command Description
Default
Parameter
Ref SCPI
[:IMMediate] Clear limit test results. √
:AUTO <b> Enable or disable clearing of limit test results
when a new trigger model cycle starts.
ON √
:AUTO? Query state of auto clear. √
:LIMit2 Limit 2 Testing: √
:UPPer Configure upper limit: √
[:DATA] <n> Specify limit; -100e6 to 100e6. 2 √
[:DATA]? Query upper limit. √
:LOWer Configure lower limit: √
[:DATA] <n> Specify limit; -100e6 to 100e6. -2 √
[:DATA]? Query lower limit. √
:STATe <b> Enable or disable Limit 2 test. OFF √
:STATe? Query state of Limit 2 test. √
:FAIL? Return result of Limit 2 test; 0 (pass) or 1(fail). √
:CLEar Clear test results: √
[:IMMediate] Clear limit test results. √
:AUTO <b> Enable or disable clearing of limit test results
when a new trigger model cycle starts.
ON √
:AUTO? Query state of auto clear. √
:IMMediate Recalculate limit tests. √
Table 14-2
CALibration command summary (user accessible)
Command Description
Default
Parameter
Ref
SCPI
:CALibration See Note.
:UNPRotected Calibration accessible to operator:
:ACALibration ACAL procedure: Sec 2
:INITiate Prepares 2182 for ACAL.
:STEP1 Performs a full ACAL (100V and 10mV).
:STEP2 Performs a limited ACAL (10mV only).
:DONE Exit ACAL mode.
:TEMPerature? Queries the internal temperature at the time of last
ACAL. If present internal temperature differs from
the last ACAL temperature by more than 1°C,
perform another ACAL.
Note: The above table only provides the commands to perform ACAL, which is a procedure to be performed by the
operator. The formal calibration is to be performed by qualified service personnel and is provided in the Model 2182
Service Manual.
Table 14-1
CALCulate command summary (cont.)