Line frequency
On power-up, the Model 2182 detects the line power frequency and automatically selects the
proper line frequency setting. The line frequency setting can be checked using the following
command:
:SYSTem:LFRequency?
The response message will be 50 or 60. The value 50 indicates that the line frequency is set
for 50Hz (or 400Hz), while 60 indicates that it is set for 60Hz.
Display
The display of the Model 2182 is primarily used to display readings, along with the units and
type of measurement. Annunciators are located at the top, bottom, left, and right of the reading
or display message. The annunciators indicate various states of operation. See “Front panel sum-
mary” (presented earlier in this section) for a complete listing of display annunciators.
NOTE The Display and Keys Test allows you to test display digit segments and annunciators,
and check the functionality of front panel keys. These tests are accessed by pressing
SHIFT and then TEST. Refer to the Model 2182 Service Manual for details.
Status and error messages
Status and error messages are displayed momentarily. During operation and programming,
you will encounter a number of front panel messages. Typical messages are either of status or
error variety, as listed in Appendix B.
Default settings
There are two default setup configurations; factory and user. As shipped from the factory, the
Model 2182 powers up to the factory default settings listed in Table 1-2. The Model 2182 can
instead be set to power up to a user default setup. The power-on default setup will be the last
configuration you saved. The SAVE key saves the present configuration as the USER power-on
setup. The RESTR key restores the instrument to the factory defaults or the user-saved defaults.
Perform the following steps to save the present setup as the power-on default configuration:
1. Configure the instrument for your measurement application.
2. Press SAVE.
3. Use the and keys to display YES or NO.
4. Press ENTER. The instrument will power-on to this USER default setup.
NOTE To assure that the proper filter state is recalled, set the analog and digital filters before
saving the user setup. See Section 3.
1-16 Getting Started

Table of Contents

Other manuals for Keithley 2182

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2182 and is the answer not in the manual?

Keithley 2182 Specifications

General IconGeneral
BrandKeithley
Model2182
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals