Additional SCPI Commands 15-3
DISPlay subsystem
The commands in this subsystem are used to control the display of the Model 2182 and are
summarized in Table 14-3.
:ENABle <b>
:DISPlay:ENABle <b> Control display circuitry
Parameters
<b> = 0 or OFF Disable display circuitry
1 or ON Enable display circuitry
Description
This command is used to enable and disable the front panel display circuitry. When disabled,
the instrument operates at a higher speed. While disabled, the display is frozen.
All front panel controls (except LOCAL) are disabled. Normal display operation can be
resumed by using the :ENABle command to enable the display or by putting the Model 2182
into local mode (press LOCAL).
:TEXT commands
:DATA <a>
:DISPlay[:WINDow[1]]:TEXT:DATA <a> Define message for display
Parameter
<a> = ASCII characters for the message (maximum of 12 characters). The characters
must be enclosed in either double quotes (“ ”) or single quotes (‘ ’).
Description
These commands define the text message for display. A message can be as long as 12
characters. A space counts as a character. Excess message characters results in an error.
:STATe <b>
:DISPlay[WINDow[1]]:TEXT:STATe <b> Control (on/off) message
Parameters
<b> = 0 or OFF Disable text message
1 or ON Enable text message
Description
This command enables and disables the text message mode. When enabled, a defined
message is displayed. When disabled, the message is removed from the display.
A user defined text message remains displayed only as long as the instrument is in remote.
Taking the instrument out of remote (by pressing the LOCAL key or sending GTL), cancels the
message and disables the text message mode.