6-8 Return to Section Topics 2600S-901-01 Rev. C / January 2008
Section 6: Range, Digits, Speed, Rel, and Filters Series 2600 System SourceMeter® Instruments Reference Manual
Rel
The rel (relative) feature can be used to null offsets or subtract a baseline reading from present
and future readings. With REL enabled, subsequent readings will be the difference between the
actual input value and the rel value as follows:
Displayed Reading = Actual Input - Rel Value
Once a rel value is established for a measurement function, the value is the same for all ranges.
For example, if 0.5A is set as a rel value on the 1A range, the rel value is also 0.5A on the lower
current ranges.
Selecting a range that cannot accommodate the rel value does not cause an overflow condition,
but it also does not increase the maximum allowable input for that range. For example, on 1A
range, the SourceMeter still overflows for a >1.02A input.
NOTE When rel is enabled, the REL annunciator turns on. Changing
measurement functions disables rel.
Front panel rel
Enabling and disabling rel
Rel can be used to null out zero offsets or to establish a zero baseline by pressing the REL key.
The reading (which becomes the rel value) is subtracted from itself. As a result, a zero reading is
displayed. Pressing REL a second time disables rel.
Defining a rel value
A unique rel value can be established for the selected measurement function from the front panel
as follows:
1. Press CONFIG then REL.
2. Choose the measurement function (CURRENT, VOLTAGE, OHMS, or WATTS), then press
ENTER or the Rotary Knob.
3. The present rel value will be displayed.
4. Set the desired rel value.
5. With the desired rel value displayed, press ENTER or the Rotary Knob, and then use EXIT
to back out of the menu structure.

Table of Contents

Other manuals for Keithley 2635

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2635 and is the answer not in the manual?

Keithley 2635 Specifications

General IconGeneral
BrandKeithley
Model2635
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals