2600S-901-01 Rev. C / January 2008 Return to Section Topics 15-3
Series 2600 System SourceMeter® Instruments Reference Manual Section 15: Performance Verification
Line power
The Series 2600 requires a line voltage of 100V to 240V and a line frequency of 50Hz or 60Hz.
Verification tests should be performed within this range.
Recommended test equipment
Table 15-1 summarizes recommended verification equipment. You can use alternate equipment as
long as that equipment has specifications equal to or greater than those listed in Table 15-1. Keep
in mind, however, that test equipment uncertainty will add to the uncertainty of each measurement.
Generally, test equipment uncertainty should be at least four times better than corresponding
Series 2600 specifications.
Table 15-1 lists the uncertainties of the recommended test equipment.
Table 15-1
Recommended verification equipment
Description Manufacturer/Model Accuracy
Digital Multimeter Agilent 3458A DC Voltage
1
(2601/2602)
DC Voltage
2
(2611/2612)
DC current
3
90mV:
0.9V:
5.4V:
36V:
190mV:
1.8V:
18V:
180V:
90nA:
0.9µA:
9µA:
90µA:
0.9mA:
9mA:
90mA:
0.9A:
±8ppm
±5ppm
±4ppm
±6ppm
±5ppm
±4ppm
±6ppm
±6ppm
±430ppm
±45ppm
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
0.5Ω , 250W, 0.1%
Precision Resistor
Isotek
RUG-Z-R500-0.1-TK3
Resistance
4
0.5Ω : ±125ppm
1GΩ , 200V, 1%
standard
Keithley
2600-STD-RES
Resistance
5
1GΩ : 250ppm
1. 90-day specifications show full-range accuracy of recommended model used for specified
measurement point.
2. Id.
3. Id.
4. Resistor used to test Model 2601/2602 3A range and Model 2611/2612/2635/2636 1.5A
range only should be characterized to uncertainty shown using resistance function of digital
multimeter before use.
5. Standard is a guarded and characterized 1 Gohm resistor used to test Model 2635/2636
100pA to 100nA current ranges.
Verification limits
The verification limits stated in this section have been calculated using only the Series 2600
one-year accuracy specifications, and they do not include test equipment uncertainty. If a
particular measurement falls outside the allowable range, recalculate new limits based both on the
Series 2600 specifications and corresponding test equipment specifications.

Table of Contents

Other manuals for Keithley 2635

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2635 and is the answer not in the manual?

Keithley 2635 Specifications

General IconGeneral
BrandKeithley
Model2635
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals