9-6 Return to Section Topics 2600S-901-01 Rev. C / January 2008
Section 9: System Expansion (TSP-Link) Series 2600 System SourceMeter® Instruments Reference Manual
An abort can also be performed by pressing the OUTPUT ON/OFF key on any node. The results
are the same as above, with the addition that all SMU outputs in the system are turned off.
Triggering with TSP-Link
TSP-Link has three synchronization lines that function similar to the Digio synchronization lines.
see “TSP-Link Synchronization lines” on page 10-10 for more information.
TSP advanced features
Use the TSP advanced features to run test scripts in parallel, to manage resources allocated to
test scripts running in parallel, and to use the data queue to facilitate real-time communication
between nodes on the TSP-Link network.
Running test scripts in parallel improves functional testing, provides higher throughput, and
expands system flexibility.
There are two methods you can use to run test scripts in parallel:
Create multiple TSP-Link networks
Use a single TSP-Link network with groups
Figure 9-2 displays the first method, which consists of multiple TSP-Link networks. Each TSP-Link
network has a master node and a GPIB connection to the PC.

Table of Contents

Other manuals for Keithley 2635

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2635 and is the answer not in the manual?

Keithley 2635 Specifications

General IconGeneral
BrandKeithley
Model2635
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals