2600S-901-01 Rev. C / January 2008 Return to Section Topics 6-11
Series 2600 System SourceMeter® Instruments Reference Manual Section 6: Range, Digits, Speed, Rel, and Filters
Filter configuration menu for Models 2635 and 2636
Press CONFIG and then FILTER to display the filter configuration menu:
• TYPE - Use this menu item to select filter type (AVERAGE or MEDIAN)
• AVERAGE - Use this menu item to select filter type (MOVING or REPEAT)
• MEDIAN - Use this menu item to select MOVING filter type.
• COUNT - Use this menu item to specify filter count (1 to 100 readings).
Enabling filter
The filter is enabled by pressing the FILTER key. The FILT annunciator is on while the filter is
enabled. Pressing FILTER a second time disables filter.
Response time
The filter parameters have speed and accuracy trade-offs for the time needed to display, store, or
output a filtered reading. These affect the number of reading conversions for speed versus
accuracy and response to input signal changes.