2600S-901-01 Rev. C / January 2008 Return to Section Topics 6-9
Series 2600 System SourceMeter® Instruments Reference Manual Section 6: Range, Digits, Speed, Rel, and Filters
Remote rel programming
Rel commands
Rel commands are summarized in Table 6-5.
Table 6-5
Rel commands
Command
1
Description
To set rel values:
smuX.measure.rel.leveli = relval
smuX.measure.rel.levelp = relval
smuX.measure.rel.levelr = relval
smuX.measure.rel.levelv = relval
To enable/disable rel:
smuX.measure.rel.enablei = smuX.REL_OFF
smuX.measure.rel.enablep = smuX.REL_OFF
smuX.measure.rel.enabler = smuX.REL_OFF
smuX.measure.rel.enablev = smuX.REL_OFF
smuX.measure.rel.enablei = smuX.REL_ON
smuX.measure.rel.enablep = smuX.REL_ON
smuX.measure.rel.enabler = smuX.REL_ON
smuX.measure.rel.enablev = smuX.REL_ON
Set current rel value.
Set power rel value.
Set resistance rel value.
Set voltage rel value.
Disable current rel.
Disable power rel.
Disable resistance rel.
Disable voltage rel.
Enable current rel.
Enable power rel.
Enable resistance rel.
Enable voltage rel.
* smuX = smua for the Model 2601/2611; smuX = smua (Channel A) or smub (Channel B) for the
Model 2602/2612.
Rel programming example
smua.reset()
--Restore Series 2600 defaults.
smua.measure.rel.leveli = 0.1
--Set current rel to 100mA.
smua.measure.rel.enablei = smua.REL_ON
--Enable current rel.

Table of Contents

Other manuals for Keithley 2635

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2635 and is the answer not in the manual?

Keithley 2635 Specifications

General IconGeneral
BrandKeithley
Model2635
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals