Page #103 background imageLoading...
Page #103 background image
Object Color Measurement
This implement is capable of simplied object color measurement through the use of the
user calibration function.
Ensure that the white calibration plate (optional accessory) and the object to be
measured are set in the same position and angle. If the illumination settings and
measurement conditions are different between the white calibration plate and object to
be measured, the measurement data is affected, resulting in incorrect measurement
data.
During the measurement, use a constant-voltage power supply for the illumination light
source to ensure conditions are as stable as possible.
Setup for Object Color Measurement
1. Prepare one or more illumination
light sources such as tungsten
lamps, and set the white
calibration plate as shown in the
gure on the right.
Set up the instrument so that it faces the
white calibration plate perpendicularly.
Set up the illumination light sources at a
45° angle.
White calibration plate
Illumination
light source
Illumination
light source
45°
45°
White Calibration
2. Perform the user calibration.
For details, refer to page 65.
Measurement
3. Set the object to be measured at the same position and angle as the
white calibration plate.
4. Measure the object.
101
Explanation and
Information
OBJECT
COLOR

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta CHROMA METER and is the answer not in the manual?

Konica Minolta CHROMA METER Specifications

General IconGeneral
BrandKonica Minolta
ModelCHROMA METER
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals