Error Messages
When the instrument does not operate properly as expected, it displays an error message
on the LCD screen. The following table shows possible error messages, their meaning
(cause), and countermeasures.
Error message Meaning (cause) Countermeasure
1
Replace the
battery.
The battery voltage is low. • Set the power switch to OFF
( side) and replace the
batteries with new ones or
use the AC adapter.
2
Input value outside
range.
The value you entered is
out of the available range.
• Enter a correct value.
Be sure to enter a value that
satises all the conditions of
the value input rule.
* Value input rule (Refer to
page 70.)
3
Specify the date
and time.
The instrument is started for
the rst time
and the date and time have
not been set.
• Set correct date and time.
4
Memory full. You attempted to save more
than 1,000 measurement
results.
• Follow the guidance and
overwrite the data or delete
unnecessary measurement
results.
5
Measurement data
not found.
You attempted to check the
measurement result or
perform target setting or
user calibration when there
is no measurement data/
stored data in the
instrument.
• Perform measurement or
select measurement data and
then try the setting again.
6
Target data does
not exist.
You attempted to set a
target when there is no
measurement data/stored
data in the instrument.
• Perform measurement or
select measurement data and
then try the setting again.
7
Backup battery has
run out.
The backup battery voltage
dropped, resulting in the
corruption of setting data.
• Charge the backup battery.
The battery is fully charged in
about 20 hours.
When the battery is fully
charged, congure the setting
of the instrument again.
• If the problem persists,
contact a KONICA MINOLTA-
authorized service facility.
103
Explanation and
Information
ERROR